Cuvette Integrity Check in Clinical Chemistry Analyzers | Accuracy & Precision - Part 8

As we have discussed in “Part 7” of these blogs, majority of fully automatic clinical chemistry analyzers having throughput of 200, 400, 800 tests / hour, are equipped with permanent cuvettes and on-board washing station. Analyzers also having many routine, special and advance wash protocols for reaction cuvette like DI water wash, special wash, detergent wash, pre-wash, post wash and wash program to reduce interference of reagents.


In high throughput analyzers, reaction cuvettes are washed and reused many times. So, it is very important to check whether reaction cuvettes are washed properly or not. As little stain on cuvette walls or any scratches will reduce accuracy & precision of test results.


As discussed in Part-4 of these blogs, Advance clinical chemistry analyzers are equipped with “Static Photometer”. This static photometer can take readings of all cuvettes with all filters in each cycle for “Cuvette Integrity Check”. i.e. 800 tests / hour clinical chemistry analyzer offered by Beacon Group is having 160 cuvettes, 12 filters and cycle time of 09 seconds. So, it can take 1920 readings (160 x 12) at every 09 seconds. 


So, advance clinical chemistry analyzers with “Static Photometer” & “Cuvette Integrity Check” ensures accurate & precise result by continuous monitoring of cuvette’s quality, after on-board washing.

  • Cuvette Integrity Check program will measure absorbance &/or transmittance of all cuvettes with all filters, continuously.
  • Monitors quality of cuvettes at each and every stage of cuvette usage. i.e. before and after using the cuvettes, before and after adding reagents / sample.
  • Check quality of each cuvette after washing and give real time status of each cuvette.
    • Cell blank checking passed - Cuvette is clean and ready to use.
    • Cell blank not passed “Error” - Cuvette is not clean, not ready to use.
  • Analyzer will not use cuvette giving “Error”, give message to user for additional or hard washing.
  • Real time status monitoring of each cuvette, at any time.
  • Just one click and window will show absorbance of each cuvette, with last 10 readings  
     

Beacon group of companies are offering range of fully automatic clinical chemistry analyzers having throughput of 200, 400 and 800 tests / hour. These clinical chemistry analyzers are equipped with all above mentioned advance features of “Cuvette Integrity Check”.